The tester has two test modes: inhalation resistance and breathing resistance
Inspiratory resistance test:
Turn on the power of the tester, switch the breathing direction to the inhalation position, press the "inhalation" button, adjust the air flow adjustment knob, when it rises to (85 ± 1) L / min, the inhalation pressure is 0 at this time. Put the sample on the head mold, adjust the wearing position of the sample and the tightness of the headband to ensure the tight fit of the mask and the head mold. Then adjust the air flow adjustment knob. When it rises to (85 ± 1) L / min, record the inspiratory resistance value at this moment, and press the "Stop" key to stop the test. During the test, take appropriate measures to avoid the specimen from attaching to the mouth of the breathing tube.
Exhalation resistance test:
Turn on the power of the tester, switch the breathing direction to the exhalation position, press the "exhalation" button, adjust the air flow knob, when it rises to (85 ± 1) L / min, the exhalation pressure is 0 at this time. Put the sample on the head mold, adjust the wearing position of the sample and the tightness of the headband to ensure the tight fit of the mask and the head mold. Then adjust the air flow adjustment knob. When it rises to (85 ± 1) L / min, record the exhalation resistance value at this moment, and press the "Stop" key to stop the test. During the test, take appropriate measures to avoid the specimen from attaching to the mouth of the breathing tube.
Technical parameter
Flow meter: (10~100)L/min,accuracy:3%;
Micro pressure sensor range:±1500Pa;accuracy:1Pa
Air supply: >100L/min;
Ventilation: (85±1)L/min;
Head model size: Big, mid, samll, size according to GB/T 32610-2016 appendix B5.5;
Display: 7” touch screen;
Print: Modular integrated thermal printer;
Communication output: RS232 (standard) / USB (optional);
Working environment: indoor temperature (20 ± 10) ℃; relative humidity <85%;
Dimensions: 420 × 600 × 720 (L × W × H)
Product features
Easy to switch between inhalation resistance and breathing resistance;
Use DGUS color touch screen as the man-machine interface;
With statistical functions, it displays the average, maximum, minimum, standard deviation and coefficient of variation.
National standard
GB 2626-2019 "Respiratory Protection Self-priming Filter Anti-Particle Respirator