概要
EMGA-920氧氮分析儀:氧氮分析儀和用戶友好產品中的新旗艦。
的EMGA-9xx系列氧氮分析儀所具備的高準確度和高重現(xiàn)性鋼鐵、新材料、催化劑等行業(yè)技術研發(fā)及品質控制的分析要求。這是按照用戶需求優(yōu)化的全新的一代產品。
氧元素以一氧化碳的形式利用非分散紅外檢測器測定。氮元素利用熱導檢測器測定。
原理及氣路圖
特征
- 超高性能
- 友好的軟件
- 全面豐富的附屬配置
- 改進的機械性能
- 樣品和助熔劑雙重投料機構(HORIBA)
規(guī)格
EMGA-920:Oxygen/Nitrogen analyzer
Principle
Oxygen:Non Dispersive Infrared detector
Nitrogen:Thermal Conductivity detector(TCD)
Measurement range*:
Oxygen:0-5%, Nitrogen:0-3%
*Up to 99% wt is possible by decreasing the sample weight.
Sample weight: 1,0 ± 0,1g
Sensitivity(Minimum reading): Oxygen/Nitrogen: 0.0000001%(m/m)
Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.000002%(m/m) or CV≦0.5% whichever is larger (Reference gas)
σn-1≦0.00003%(m/m) or CV≦1.0% whichever is larger (Standard sample)
** Based on the sample weight 1,0 ± 0,1g and with automation functions described below
Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD
Power of furnace
Iner gas fusion in impulse furnace
Enable to set the generating power 0 - 8,0kw
Sample introduction: Dual sample/flux introduction mechanism
Automation functions: Auto cleaner, Crucible loader
Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter
Sample ID: Enter up to 20 one-byte characters
Calibration
1) One point or multi point calibration (Reference gas or standard samples)
2) Calibration using previous analysis data
3) Calibration curve correction function
Functions
1) Display of realtime extraction curve
2) Analysis interruption
3) Self diagnosis and alarm display
4) Analysis of extraction curve (Superposition, difference etc analysis)
5) Out put (RC-232C or TCP/IP)
Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.
Weight: 230kg
On carrying, it is devided in two units (Under 140kg/unit)
Computer: PC model operating OS Windows® XP(SP2) and beyond
Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz
Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1,5kVA (MAX)
Ground resistance: Less than 10Ω
Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, Linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band
Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe(O.D.6mm) and suitable connecter fitting within 5m of unit
Cooling mechanism: Radiator unit (Separation type)
Electronic balance (Option)
Enable connection with electronic balance with 1-0,01mg minimum display
Optional automatic voltage regulator (AVR)
Capacity:15kVA